The following links will give you more information about the PowerProMax™ System and it's Sterilization Parameters

Handpiece Sterilization Parameters
SureCharge® Sterilization Parameters
Sterile Transfer System Sterilization Parameters
Sterilization Parameters for Standard Batteries (NiCad)

Handpieces
Small Bone
Batteries
Battery Charger
Sterile Transfer Battery System
SureCharge® System
Accessories
Attachments
Knee Blades
Saw Blade Guide

 

Handpiece Sterilization Parameters

Sterilization Parameters for sterilizing PowerProMax Battery, Electric II, and Pneumatic Handpieces and attachments with or without the PowerPro System Sterilization Case PRO5090 or PRO5095.

Sterilization Type Temperature Exposure Time Dry Time
Steam Pre-vacuum
Steam Gravity
Steam Gravity
 
270 - 272° F (132 - 133°C)
270 - 272° F (132 - 133°C)
250 - 254° F (121 - 123°C)
4 minutes
10 minutes
40 minutes
8 minutes minimum*
8 minutes minimum*
8 minutes minimum*
 

*Caution: An eight (8) minute minimum dry cycle is strongly recommended on all handpieces and attachments every time the product is sterilized. Failure to use a dry cycle on the products may lead to reduced product performance or premature product failure. Operation of a handpiece that is not completely cool and dry may decrease performance and /or reliability.
Steris is not an approved sterilization method for any Hall® Products.

SureCharge® Sterilization Parameters

Sterilization Parameters for sterilizing PowerPro Battery SureCharge Sterilization Charging Case (PRO3200).

Sterilization Type Temperature Exposure Time Dry Time
Steam Pre-vacuum
 
270 - 272° F (132 - 133°C) 3 minutes 8 minutes minimum*
 

*Caution: The SureCharge sterilization case may not be sterilized using a steam gravity cycle. Prior to sterilizing the SureCharge System always discharge batteries and insert a new PRO3205 disposable filter. An eight (8) minute minimum dry cycle must be used when sterilizing the PRO3200 SureCharge Sterilization Case. Failure to use the minimum dry cycle may result in moisture left in the container.
Steris is not an approved sterilization method for any PowerPro products.

Sterile Transfer System Sterilization Parameters

Sterilization parameters for PowerPro Sterile Transfer Battery Case (PRO3110) and PowerPro Sterile Transfer Shroud (PRO3130).

Sterilization Type Temperature Exposure Time Dry Time
Steam Pre-vacuum
Steam Gravity
 
270 - 272° F (132 - 133°C)
250 - 254° F (121 - 123°C)
3 minutes
10 minutes
8 minutes minimum*
8 minutes minimum*
 

*Caution: Do not sterilize PRO3115 Battery Pack (NiMH) or PRO3135 Sterile Transfer Platform.
Steris is not an approved sterilization method for any PowerPro products.

Sterilization Parameters for Standard Batteries (NiCad)

Sterilization parameters for PowerPro Standard Large (PRO3010) and Small (PRO3020) Battery Packs.

Sterilization Type Temperature Exposure Time Dry Time
Steam Pre-vacuum
Steam Gravity
 
270 - 272° F (132 - 133°C)
250 - 254° F (121 - 123°C)
3 minutes
3 minutes
None
None
 

*Caution: Do not leave batteries in the autoclave longer than the recommended sterilization exposure time. Damage may result, thus shortening the useful life. PowerPro Battery Packs (PRO3010 and PRO3020) are qualified for use in STERRAD Sterilization Systems.
Steris is not an approved sterilization method for any PowerPro products.

 